- Turn on pH meter.
- Remove the electrode cap by gently unscrewing the white neck of the cap.
- Place the cap on the table so that it is not in the way.
- Place the pH rinse beaker under the pH probe.
- Thoroughly rinse the entire black section of the probe with dH2O.
- Dry the probe gently with a paper towel. Be sure to gently dab the tip of the probe.
- Position the black stand beneath the probe of the pH meter.
- Place the pH 1 standard on the stage.
- Press the calibrate button.
- Lower the arm of the pH meter so that the probe is completely submerged in the buffer solution. Your setup should look like Figure 2a at this point.
- Wait until the pH meter is done calibrating. This is indicated by a radical symbol on the M on the top right of the pH meter screen. See Figure 2b.
- Click the read button.
- Raise the arm so the probe is no longer in the buffer solution.
- Repeat steps 4-12 with the other buffer solutions.
Tip
It may be helpful to view the following video for additional instruction (see below): https://www.youtube.com/watch?v=7yuXjGdQRzM
(a)
|
(b)
|
Figure 2. (a) pH meter during calibration. (b) pH meter screen when it is ready to read (notice the M in the top right). |
|